Calibration is a critical maintenance step that helps maintain the scanning accuracy of an intraoral scanner. Performing it at the right intervals and following the correct procedure reduces the risk of inaccurate impressions and clinical remakes.
Recommended Calibration Frequency
| Scenario | Recommended Frequency |
|---|---|
| Routine daily use | Every 2 weeks |
| High-volume use (10+ cases per day) | Once a week |
| After moving the scanner to a different room | Immediately |
| After the cart is bumped or subjected to shock | Immediately |
| Significant ambient temperature change (>5 °C) | Immediately |
| Suspected accuracy issues (poor restoration fit) | Immediately |
| After changing the scanning tip | Recommended |
| After major software/firmware update | Recommended |
| After long-term storage (>1 month) before reuse | Immediately |
Some newer systems (e.g., certain 3Shape TRIOS models) are factory-pre-calibrated and claim reduced need for user calibration. Even so, periodic verification with a reference object remains good practice.
General Calibration Procedure
Power on the scanner and allow it to warm up for 10–15 minutes so the optical components reach thermal stability.
Perform calibration in a stable room-temperature environment (ideally 20–25 °C), away from direct air-conditioning airflow.
Clean the tip lens thoroughly and confirm there are no stains or scratches.
Prepare the official calibration tool (calibrator). Ensure it is clean and has not been dropped or damaged.
Place the calibrator on a stable, level surface.
Enter calibration mode in the software and follow the on-screen prompts.
Insert the scanner tip correctly and firmly into the calibrator slot; keep the device steady.
Wait for the process to complete (typically 30 seconds to a few minutes). Do not touch the equipment during this time.
Confirm the "Calibration successful" message before removing the tip.
If calibration fails, clean the lens and calibrator and retry. Contact technical support if failures continue.
Always follow the exact steps and prompts shown in your scanner's software and user manual.
Care of the Calibration Tool
The calibrator is a precision reference object containing accurately dimensioned markers.
Store it in its protective dust bag when not in use; protect it from dust and impact.
Wipe the exterior with 70–75 % ethanol gauze. Do not open the housing.
Dropping the calibrator can shift internal markers and require factory re-certification.
If lost or damaged, purchase a genuine replacement from the manufacturer (typically several hundred dollars).
Common Causes of Calibration Failure
Dirty or scratched tip lens
Dirty or damaged calibrator
Incorrect seating of the tip in the calibrator
Excessive ambient light interference
Scanner not yet thermally stable
Tip has reached the end of its service life
Internal optical misalignment (requires factory service)
Regular calibration, combined with proper tip cleaning and environmental control, helps preserve the accuracy of your digital impressions. For the Aident AI-30 series, follow the specific calibration instructions provided in the user manual and software. Genuine calibration tools and support are available through official channels.
For detailed operating manuals, calibration guidance, or to order genuine accessories for the AI-30, visit:
https://www.aident3d.com/3d-scanner/intraoral-scanner/
or contact the team at https://www.aident3d.com/inquiry / https://www.aident3d.com/download
Consistent calibration is one of the simplest ways to protect clinical outcomes and the long-term performance of your intraoral scanner.