Intraoral Scanner Calibration Frequency and Method: How Often Should You Calibrate?

Aug 30, 2026

Leave a message

Calibration is a critical maintenance step that helps maintain the scanning accuracy of an intraoral scanner. Performing it at the right intervals and following the correct procedure reduces the risk of inaccurate impressions and clinical remakes.

Recommended Calibration Frequency

Scenario Recommended Frequency
Routine daily use Every 2 weeks
High-volume use (10+ cases per day) Once a week
After moving the scanner to a different room Immediately
After the cart is bumped or subjected to shock Immediately
Significant ambient temperature change (>5 °C) Immediately
Suspected accuracy issues (poor restoration fit) Immediately
After changing the scanning tip Recommended
After major software/firmware update Recommended
After long-term storage (>1 month) before reuse Immediately

Some newer systems (e.g., certain 3Shape TRIOS models) are factory-pre-calibrated and claim reduced need for user calibration. Even so, periodic verification with a reference object remains good practice.

General Calibration Procedure

Power on the scanner and allow it to warm up for 10–15 minutes so the optical components reach thermal stability.

Perform calibration in a stable room-temperature environment (ideally 20–25 °C), away from direct air-conditioning airflow.

Clean the tip lens thoroughly and confirm there are no stains or scratches.

Prepare the official calibration tool (calibrator). Ensure it is clean and has not been dropped or damaged.

Place the calibrator on a stable, level surface.

Enter calibration mode in the software and follow the on-screen prompts.

Insert the scanner tip correctly and firmly into the calibrator slot; keep the device steady.

Wait for the process to complete (typically 30 seconds to a few minutes). Do not touch the equipment during this time.

Confirm the "Calibration successful" message before removing the tip.

If calibration fails, clean the lens and calibrator and retry. Contact technical support if failures continue.

Always follow the exact steps and prompts shown in your scanner's software and user manual.

Care of the Calibration Tool

The calibrator is a precision reference object containing accurately dimensioned markers.

Store it in its protective dust bag when not in use; protect it from dust and impact.

Wipe the exterior with 70–75 % ethanol gauze. Do not open the housing.

Dropping the calibrator can shift internal markers and require factory re-certification.

If lost or damaged, purchase a genuine replacement from the manufacturer (typically several hundred dollars).

Common Causes of Calibration Failure

Dirty or scratched tip lens

Dirty or damaged calibrator

Incorrect seating of the tip in the calibrator

Excessive ambient light interference

Scanner not yet thermally stable

Tip has reached the end of its service life

Internal optical misalignment (requires factory service)

Regular calibration, combined with proper tip cleaning and environmental control, helps preserve the accuracy of your digital impressions. For the Aident AI-30 series, follow the specific calibration instructions provided in the user manual and software. Genuine calibration tools and support are available through official channels.

For detailed operating manuals, calibration guidance, or to order genuine accessories for the AI-30, visit:
https://www.aident3d.com/3d-scanner/intraoral-scanner/
or contact the team at https://www.aident3d.com/inquiry / https://www.aident3d.com/download

Consistent calibration is one of the simplest ways to protect clinical outcomes and the long-term performance of your intraoral scanner.